MS-2 CCD Reader User’s Manual 7-5
Utilities
9. Click Open, select the appropriate firmware file (a .mhx file) and click OK.
10. Click the Set Environment button to bring up the COM Port menu.
11. Select the appropriate COM Port and click OK.
12. Click the Download button. When the following dialog appears, cycle power to the
reader (turn power off and then back on) and then click OK to continue.
13. Click OK again once firmware is downloaded.
14. Click the Erase button to clear old firmware.
15. Click OK again when you see the dialog stating that old firmware has been erased.
16. Click the Full Operation(D+E+B+P) button. When the dialog shown in step 12
appears, cycle power to the reader (turn power off and then back on), and then click
OK to complete the download process.
17. Click OK again when you see the dialog stating that the firmware download is complete.
18. Detach the
Firmware Upgrade Adapter (61-000143-01) and re-connect the MS-2 in its
original configuration.
19. Verify installation of the new firmware using Firmware Verification in ESP Utilities.
20. If the firmware update procedure fails, check hardware and pinout connectivity, and
repeat the procedure.
Caution: Do not interrupt power or disconnect the host cable while download is in
progress. Be sure that each download is complete before moving on to the next.
Important: When updating firmware, be sure that the application code and boot code
versions are mutually compatible.