MINI Hawk Xi High Performance Imager User Manual
Calibration Options
Symbol Type
Disabled
When this feature is disabled, only the currently-enabled symbologies will be considered
during the calibration process.
Enabled
When this feature is enabled, autodiscrimination will be in effect during the calibration
process. All symbologies that are readable by the imager except PDF417 and Pharmacode
will be considered during calibration. Any new symbologies successfully decoded during
calibration will remain enabled at the end of the process. All enabled symbologies will
remain enabled.
For example, assume that only Code 39 is enabled at the beginning of calibration. If a
Code 128 symbol is read during calibration, then Code 128, as well as Code 39, will be
enabled after the calibration process is complete.
Serial Cmd: <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI
margin,line scan height,processing>
Default: Enabled
Options: 0 = Disabled 1 = Enabled