Quadrus EZ Reader User’s Manual 6-11
Symbol Quality
Cell Fill
Definition:
Percentage of the ideal cell size that the dot fills. Worst Case indicates the
quality of the worst dot in percentage and its position in the grid.
For dot peen symbols this is referred to as Cell Size, and for Laser or
Chemical this is called Cell Fill. The calculation used for both is similar.
If enabled, the grading letter is appended to the symbol.
Grading:
A if 0 - 2%
B if 3 - 7%
C if 8 - 13%
D if 14 - 20%
F if > 20%
Dot Peen, Round Laser/Chemical measures dot size from center of the dot,
averages the maximum and minimum, and compares it with the ideal cell
size to find the percentage.
Square Laser/Chemical measures dot size from left-right and up-down, on
the angle of the symbol, averages the dimensions, and compares it with the
ideal cell size to find the percentage.
Serial Cmd: <K712,dot center offset,cell fill,dot ovality,angle of distortion>
Default: Disabled
Options:
0 = Disabled 1 = Worst Case
2 = Average 3 = Both