Quadrus EZ Reader User’s Manual 7-11
Matchcode
Sequence Step
As an example of Sequence Step, if Sequence Step is set to 3 and Sequential Matching
is set to Increment.
Usage:
Useful in applications in which it is desirable to count by intervals other than 1.
Definition:
Sequencing in Matchcode operations can occur in steps from 1 to 32,768.
Sequencing performs like a mechanical counter by displaying positive inte-
gers and a specific number of digits after roll-overs. For example, 000 – 3 =
997 (not –3) and 999 + 3 = 002 (not 1002).
Serial Cmd: <K228,sequence step>
Default: 1
Options: 1 to 32,768
Master symbol Decoded symbol Master symbol after decode
003 001 003
003 002 003
003 003 006
006 004 006
006 005 006
006 006 009