Chapter 6 Autocalibration
MS-880 Industrial Long Range Scanner User’s Manual 6-3
6–Autocalibration
Autocalibration Steps
Follow this procedure for all three methods of autocalibration:
1. Set your scan rate at the appropriate speed for your application.
2. Position your bar code symbol at the distance used in your application and confirm
that it is at a distance within the specifications for that symbol size (see “Position
Scanner and Symbol” on page 1-12).
3. If you are not certain of the code type, enable all code types by sending a <P>
command from the terminal window, or clicking on the Autodiscrimination but-
ton in Calibration/Advanced in ESP.
1
4. Start the autocalibration procedure.
Note: Once a successful autocalibration is completed, you may want to save
those setting to the calibration database (see Chapter 7, “Calibration Data-
base”).
5. Repeat autocalibration at other ranges and/or label types you might be scanning
in your application.
6. Check your readings at each
end of the scan width
required for your applica-
tion, keeping in mind that
the scan width testing may
not produce results that are
entirely symmetrical.
Note: If you are setting up
for Laser Framing (see
page 9-6), be sure that you
only test for autocalibration
in the active areas of the
scan beam.
7. To save the new settings
(including focus, gain, and
tracking) for power-up,
send a <Zp> command.
Note: To silence the beeper, send <K702,0> or remove or block the test label.
After autocalibration has passed, the scanner will beep once, automatically exit the
autocalibration routine and begin scanning for a label in Continuous Read mode,
beeping for each good read.
1. Note: If using an I 2 of 5 label, verify that the number of characters in the label being
scanned matches the code length enabled for the I 2 of 5 code type (default is 10 and 6). See
“Interleaved 2 of 5” on page 5-14.
Figure 6-2 Autocalibration Setup