Autocalibration by Embedded Menu
6-10 MS-880 Industrial Long Range Scanner User’s Manual
6–Autocalibration
Autocalibration Menu Options
START Autocalibration
Typing A initiates an Autocalibration sequence. Once entered, no other keystrokes
will be accepted with the exception of the ESC key, which terminates the routine.
RECALL Initial Settings
Typing B restores the original settings that were present upon entering the Autocali-
bration Setup Menu. These settings include Motor Speed, Gain, Tracking, and
Focus.
SAVE Settings to Database
Typing C saves the current Gain, Tracking, and Focus settings to the Configura-
tion Database. The database is made up of indexes 1 through 7. Current settings will
be saved to the first open index. See Chapter 8, “Optoelectrics.”
EXIT Calibration Setup menu
To exit the Autocalibration Setup menu, press the ESC key. The following prompt
will be displayed:
“Do you want to save calibration settings for power on? (Y=yes N=no).”
Typing Y will cause the current settings in the Calibration Setup menu to be saved
to Flash Memory for recall when the scanner is powered up at a later date. These set-
tings include Motor Speed, Gain, Tracking, Focus, AGC Status, Laser Power, and
Raster Status. Entering N will cause the scanner to be reset without saving the set-
tings to Flash memory.
Menu Selections
Typing 1 will take you to the Calibration Results menu (page 6-11).
Typing 2 will take you to the Raster/Laser Framing Setup menu (page 9-9).
Typing 3 will take you to the Configuration Database menu (page 7-6).
Typing 4 will take you to the Advanced Setup menu (page 6-12).