Microscan MS-4X Manual de usuario Pagina 152

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 312
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 151
Laser Setup
9-8 MS-880 Industrial Long Range Scanner User’s Manual
9–Raster/Laser Con-9–Raster/Laser Con-
Laser Off Time
When setting up Laser Framing in ESP-MP, use the Off Time and On Time arrows to
constrain or expand the laser framing area.
Laser On Time
Laser Power
Definition: Duration of laser off time, in increments of 1/255th of the total scan
width.
Important: The combined values of Laser Off Time and Laser On
Time cannot exceed 255 which is the duration of one scan.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: 0
Options: 0 to 255
Definition: Duration of laser on time, in increments of 1/255th of the total scan
width.
Important: The combined values of Laser Off Time and Laser On
Time cannot exceed 255 which is the duration of one scan.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: 255
Options: 0 to 255
Definition: The user can select the Low or High laser power.
Usage: Applications that read past 45 inches typically use High.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: Low
Options: 0 = Low (laser power = 2 mW)
1 = High (laser power = 3.25 mW)
Vista de pagina 151
1 2 ... 147 148 149 150 151 152 153 154 155 156 157 ... 311 312

Comentarios a estos manuales

Sin comentarios