EZ Match User Manual
4-35
Communications
Postamble
Postamble Status
Postamble Characters
Important:
The ASCII characters
<
,
>
, and
,
can only be entered as hex pairs:
For <: <K142h,,3C>
For >: <K142h,,3E>
For
, : <K142h,,2C>
See ASCII Character Entry Modifier on page 4-33 for a detailed explanation of how to
enter ASCII characters as hex values.
Usage: Useful for identifying and controlling incoming data. For example, defining
the postamble as a carriage return and a line feed causes each decoded
message to be displayed on its own line.
Definition: Allows the user to enable or disable up to four postamble characters that
can be added to the end of the decoded data.
Serial Cmd: <K142,status,postamble character(s)>
Default: Enabled
Options: 0 = Disabled 1 = Enabled (within any protocol)
Serial Cmd: <K142,status,postamble character(s)>
Default: ^M^J corresponds to: carriage return/line feed.
Options: To enter control characters within a serial command, hold down the control
key while typing the desired character.
Example: <K142,1,CNTL-m CNTL-j> to enter ^M^J.