3-18
EZ Match User Manual
Tube/Cap Information
Tube Angle
Notes:
• This value is not output in Read Rate results.
•The Tube Angle value is not required to match when requiring more than
one tube/cap to qualify for an output.
Tube/Cap Tip Position
Cap Color (EZ Match Color Only)
Definition When Tube Angle is enabled, a polarity sign (+/–) and a three-digit value
representing the angle of the tube in respect to the field of view will be
appended to tube/cap detection output as a preamble.
A + value indicates that the tube is angled away from the center and toward
the bottom of the field of view, while a – indicates that the tube is angled
away from the center and toward the top of the field of view.
Serial Cmd: <K850,separator,tube/cap library index,tube angle
,tube/cap tip position,cap
color,match status,tube diameter
>
Default: Disabled
Options: 0 = Disabled
1 = Enabled
Definition When Tube/Cap Tip Position is enabled, a four-digit value will be
appended to tube/cap detection output as a preamble.
This value represents the distance on the X-axis from the left of the captured
image to the tip of the tube or cap, in millimeters x10.
Example: 0835 represents 83.5 mm.
Serial Cmd: <K850,separator,tube/cap library index,tube angle
,
tube/cap tip position
,
cap color,match status,tube diameter
>
Default: Disabled
Options: 0 = Disabled
1 = Enabled
Definition This feature is only available for the color version of the EZ Match Imager.
Serial Cmd: <K850,separator,tube/cap library index,tube angle
,tube/cap tip position,
cap
color
,match status,tube diameter
>
Default: Disabled
Options: 0 = Disabled 1 = Color 2 = HSV