EZ Match User Manual
Multi-Operation by Serial Command
Operation (Multi-Operation)
Disabled
When set to Disabled, no processing is specified for the operation.
Bar Code Reader
Read cycle operation for the duration of the specified time and number of frames will
include processing of image frames for enabled symbologies.
Tube Detection
Read cycle operation for the duration of the specified time and number of frames will
include processing of image frames for presence or absence of test tubes.
Cap Detection
Read cycle operation for the duration of the specified time and number of frames will
include processing of image frames for presence or absence of test tube caps.
Frame Count (Multi-Operation)
Definition: Specifies the operational flow of the read cycle, and allows up to 3 different
operations to be sequenced in a single read cycle. The operation field is a
bit field, and multiple processes can be defined for a single operation. The
examples below demonstrate how to configure a tube and cap detection
process to be defined in a single operation.
Tube and Cap Detection Operation = 2 + 4 = 6
Serial Cmd: <K260,index,operation,frame count,timeout,database min,database max>
Options: Disabled = 0
Bar Code Reader = Add 1
Tube Detection = Add 2
Cap Detection = Add 4
Definition: Frame Count only applies if the imager is configured in a Rapid Capture
mode of operation. The frame count value indicates the number of image
frames that will occur before the read cycle will switch to the next operation.
Serial Cmd: <K260,index,operation,frame count,timeout,database min,database max>
Default: 0 = Disabled
Options: 1 to 6